Magnetic Susceptibility Instrument for Magnetically Inhomogeneous Granular Mixtures

dc.contributor.authorCharles T. Pett
dc.date.accessioned2024-02-01T14:17:34Z
dc.date.available2024-02-01T14:17:34Z
dc.date.issued2024
dc.descriptionA Digilent Analogue Discovery 2 (AD2) was used in conjunction with the Digilent WaveForm software Network Analyzer to find the resonance frequency of our LRC circuit. The resonnance frequency was used to calculate inductance. We used the measured inductance to calculate the permeability of granular mixture relative to air to solve for the magnetic susceptibility of the mixture.
dc.description.abstractWe introduce an instrument and novel method for characterizing the bulk magnetic susceptibility of granular mixtures by submerging an inductor coil in a bed of metallic beads and gauging changes in self-inductance. The resonance frequency of the coil was measured to determine its inductance and evaluate the magnetic permeability of ferrous mixtures relative to air. In air, our coil was accurate to within 0.1\% of the permeability of free space. The range of magnetic susceptibility values for magnetically inhomogeneous granular mixtures is poorly constrained, but our coil uniquely quantifies bulk effects that other surface meters are not designed to resolve. Compared to both a commercial Terraplus Inc. KT-10 meter and theoretical approximations, we report similar trends in susceptibility values measured as a function of mass of ferromagnetic material per volume.
dc.identifierhttps://doi.org/10.13016/dspace/h9sj-oasd
dc.identifier.urihttp://hdl.handle.net/1903/31634
dc.language.isoen_US
dc.publisherReview of Scientific Instruments
dc.relation.isAvailableAtDigital Repository at the University of Maryland
dc.relation.isAvailableAtUniversity of Maryland (College Park, Md)
dc.rightsCC0 1.0 Universal*
dc.rights.urihttp://creativecommons.org/publicdomain/zero/1.0/*
dc.subjectmagnetic susceptibility
dc.subjectscientific instrument
dc.subjectgranular matter
dc.titleMagnetic Susceptibility Instrument for Magnetically Inhomogeneous Granular Mixtures
dc.typeDataset
local.equitableAccessSubmissionYes

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
susceptibility measurements.zip
Size:
1.15 MB
Format:
Unknown data format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.55 KB
Format:
Item-specific license agreed upon to submission
Description: