3D Magnetic Imaging using SQUIDs and Spin-valve Sensors

Loading...
Thumbnail Image

Publication or External Link

Date

2016

Citation

Abstract

We have used 2 µm by 4 µm thin-film Cu-Mn-Ir spin-valve sensors and high Tc YBa2Cu3O7-x dc SQUIDs to take magnetic images of test samples with current paths that meander between 1 and 5 metallization layers separated by 1 µm to 10 µm vertically. I describe the development and performance of a 3D magnetic inverse for reconstructing current paths from a magnetic image. I present results from this inverse technique that demonstrate the reconstruction of the 3D current paths from magnetic images of samples. This technique not only maps active current paths in the sample but also extracts key parameters such as the layer-to-layer separations. When imaging with 2 µm by 4 µm spin-valve sensors I typically applied currents of 1 mA at 95 kHz and achieved system noise of about 200 nT for a 3 ms averaging time per pixel. This enabled a vertical resolution of 1 µm and a lateral resolution of 1 µm in the top layers and 3 µm in the bottom layer. For our roughly 30 µm square SQUID sensors, I typically applied currents of 1 mA at 5.3 kHz, and achieved system noise of about 200 pT for a 3 ms averaging time per pixel. The higher sensitivity compared to the spin-valve sensor allowed me to resolve more deeply buried current paths.