Humeral Fracture Fixation Techniques: A FEA comparison of locing and compression techniques with cadaveric pullout comparison of cortical compression and internal locking screws.

dc.contributor.advisorBarker, Donalden_US
dc.contributor.authorJohnson, Aaronen_US
dc.contributor.departmentMechanical Engineeringen_US
dc.contributor.publisherDigital Repository at the University of Marylanden_US
dc.contributor.publisherUniversity of Maryland (College Park, Md.)en_US
dc.date.accessioned2007-09-28T15:02:17Z
dc.date.available2007-09-28T15:02:17Z
dc.date.issued2007-08-13en_US
dc.description.abstractLocking and non-locking humeral repair techniques provide different mechanical constructs for securing fractures, and consequently could generate different strain fields at the callus. The purpose of this study was to investigate the strain field callus, and to compare to determine if one construct offers a healing advantage over another. An FEA analysis was conducted using ABAQUS, with all contact surfaces modeled as friction interfaces; additionally, a pretension was applied to the non-locking construct to simulate the effect of installation. The models were subjected to axial tension loads, and results were compared with existing cadaveric and synthetic experimental loading. Additional validation involved screw pullout testing conducted on cadaveric humeri. Results showed that the strain fields at the fracture site showed no significant variation in distribution, shape, or magnitude, therefore concluding that the locking plate offered no biomechanical healing advantage.en_US
dc.format.extent810982 bytes
dc.format.mimetypeapplication/pdf
dc.identifier.urihttp://hdl.handle.net/1903/7367
dc.language.isoen_US
dc.subject.pqcontrolledEngineering, Mechanicalen_US
dc.subject.pqcontrolledHealth Sciences, Medicine and Surgeryen_US
dc.subject.pquncontrolledhumeral FEA comparisonen_US
dc.subject.pquncontrolledlocking and cortical screw pullouten_US
dc.subject.pquncontrolledhumeral fixationen_US
dc.subject.pquncontrolledorthopedicen_US
dc.subject.pquncontrolledbiomechanicsen_US
dc.titleHumeral Fracture Fixation Techniques: A FEA comparison of locing and compression techniques with cadaveric pullout comparison of cortical compression and internal locking screws.en_US